Source door open error message on Xevo TQ MS - WKB76152
Article number: 76152
SYMPTOMS
- Source door open error message mid sample set
- Replaced these parts with no improvement:
- Enclosure interlock cable assembly (700004348)
- Interlock cable (700008862)
- Pressure sensor PCB (700004196)
- Pneumatic pressure PCB (700011435)
- Source enclosure (700004319)
- Source PCB (700010022)
ENVIRONMENT
- Xevo TQ MS
CAUSE
Faulty trap bottle
FIX or WORKAROUND
Replace the trap bottle (700004606).
ADDITIONAL INFORMATION
id76152, XEVOTQ, XEVOTQA, XEVOTQIVD