BioAccord LockSpray check and injection errors - WKB75317
Article number: 75317
SYMPTOMS
- On completion of the run there is a LockSpray "Check Error" message in the Health Center for the ACQUITY RDa and the analysis says "Injection Error"
- Following the health checks, purging the fluidics clears the error but the analysis is still stuck in an injection error
- Resubmitting samples results in a second LockSpray collection error
ENVIRONMENT
- ACQUITY RDa
- BioAccord
CAUSE
Insufficient LockSpray signal due to a dirty ion block and ion guide assembly
FIX or WORKAROUND
- Replace the aperture disk.
- Replace ESI capillary probe.
- Clean the ion block and ion guide assembly.
ADDITIONAL INFORMATION
