BioAccord LockSpray check and injection errors - WKB75317

Article number: 75317

SYMPTOMS

  • On completion of the run there is a LockSpray "Check Error" message in the Health Center for the ACQUITY RDa and the analysis says "Injection Error"
  • Following the health checks, purging the fluidics clears the error but the analysis is still stuck in an injection error
  • Resubmitting samples results in a second LockSpray collection error

ENVIRONMENT

  • ACQUITY RDa
  • BioAccord

CAUSE

Insufficient LockSpray signal due to a dirty ion block and ion guide assembly

FIX or WORKAROUND

  1. Replace the aperture disk.
  2. Replace ESI capillary probe.
  3. Clean the ion block and ion guide assembly.

ADDITIONAL INFORMATION

 