How to change from LockSpray to MALDI source and from MALDI source to LockSpray? - WKB74747

Article number: 74747

OBJECTIVE or GOAL

  • Change from LockSpray to MALDI source.
  • Change from MALDI source to LockSpray.

ENVIRONMENT

  • SYNAPT G2
  • SYNAPT G2-S
  • SYNAPT G2-Si
  • SYNAPT HDMS
  • MALDI
  • LockSpray

PROCEDURE

From LockSpray to MALDI:
1. On the Tune page, click Vacuum > Vent to vent the instrument.
2. Close the MassLynx software.
3. Change to MALDI.
4. Start up MassLynx.
* When MassLynx starts up, a window appears indicating that you have changed to the MALDI source.
5. On the Tune page, click Vacuum > Pump to pump down.

From MALDI to LockSpray:
1. On the Tune page, click Vacuum > Vent to vent the instrument.
2. Close the MassLynx software.
3. Change to LockSpray.
4. Start up MassLynx.
* When MassLynx starts up, a window appears indicating that you have changed to the ESCi source.

ADDITIONAL INFORMATION

If you have any problems switching sources, try the following:

  1. Close MassLynx.
  2. Run "stop acquity processes".
  3. Remove the two color-coded source cables.
  4. Perform electronic reboot and wait five minutes.
  5. Open MassLynx
  6. Replace the two color-coded cables
  7. Accept source recognition. 

You can also check the ACQUITY console window drop-down menu for the source type being used and select ESI or MALDI depending on the source currently connected in order to force MassLynx to recognize the new source.

A PC reboot may also be required.

 

 

id74747, Electrospray, MALDIG2SI, MALDIG2SI, MALDIMICRO, SYNAPT, SYNAPTMS, SYNG2HD, SYNG2HDMAL, SYNG2MS, SYNG2MSMAL, SYNG2SHD, SYNG2SHDML, SYNG2SIHD, SYNG2SIMS, SYNG2SMS, SYNG2SMSML