Xevo QTof system vents after sample cone is replaced - WKB74110
Article number: 74110
SYMPTOMS
- After sample cone is cleaned, the system vents
ENVIRONMENT
- Xevo G2-XS
- UNIFI
- Xevo G3 QTof
- Cyclic IMS
CAUSE
Poorly seated sample cone
FIX or WORKAROUND
- Remove the sample cone and refit it.
- Select Pump in the software.
ADDITIONAL INFORMATION
id74110, XEVOG2XSTF