Skip to main content
Waters

Xevo QTof system vents after sample cone is replaced - WKB74110

Article number: 74110

SYMPTOMS

  • After sample cone is cleaned, the system vents

ENVIRONMENT

  • Xevo G2-XS
  • UNIFI
  • Xevo G3 QTof
  • Cyclic IMS

CAUSE

Poorly seated sample cone

FIX or WORKAROUND

  1. Remove the sample cone and refit it.
  2. Select Pump in the software.

ADDITIONAL INFORMATION

Not able to find a solution? Click here to request help.