Skip to main content
Waters

High contaminant ion persists on SYNAPT G2-Si after lots of flushing and cleaning the ion block - WKB72301

Article number: 72301

SYMPTOMS

  • e6 counts remain for the contaminant ion even when infusing from an external pump straight to the probe

ENVIRONMENT

  • SYNAPT G2-Si
  • All MS

CAUSE

The source enclosure and pumping block need to be cleaned.

FIX or WORKAROUND

Clean the source enclosure and pumping block with a suitable solvent (for example, methanol/water or IPA/water).

ADDITIONAL INFORMATION

Users should avoid getting their gloves wet with solvent while cleaning; it could result in PEG contamination.

id72301, A-SQ, isopropanol, MALDIG2SI, SQD2, SYNAPT, SYNAPTMS, SYNG2HD, SYNG2HDMAL, SYNG2MS, SYNG2MSMAL, SYNG2SHD, SYNG2SHDML, SYNG2SIHD, SYNG2SIMS, SYNG2SMS, SYNG2SMSML, XEVOG2QTOF, XEVOG2SQTF, XEVOG2STOF, XEVOG2TOF, XEVOG2XSQT, XEVOQTOF, XEVOTQ, XEVOTQA, XEVOTQD, XEVOTQDIVD, XEVOTQIVD, XEVOTQS, XEVOTQSIVD, XEVOTQSMIC, XEVOTQXS, XEVTQSMIVD, XEVTQXSIVD

Not able to find a solution? Click here to request help.