Random source exhaust warning messages on the Xevo TQD - WKB6964

Article number: 6964

SYMPTOMS

  • Random source exhaust warning messages in the early morning hours
  • Tubing from the MS to the trap bottle has a loop
  • Cannot position trap bottle to eliminate loop
  • Condensation appears in the tubing to the trap bottle
  • Tubing from the trap bottle to exhaust goes straight up and has no condensation

ENVIRONMENT

  • Xevo TQD

CAUSE

The room temperature in the lab drops overnight, causing condensation in the tubing loop to the trap bottle.

FIX or WORKAROUND

Increase the lab room temperature slightly.

ADDITIONAL INFORMATION