Poor resolution and peak shape on SYNAPT G2-Si - WKB69286

Article number: 69286

SYMPTOMS

  • Leucine enkephalin peak resolution <6000; peak shape very poor
  • SYNAPT moved to new lab and onto a new computer
  • SYNAPT was powered-off for ~2 weeks during move
  • No serial number listed in Target Registry

ENVIRONMENT

  • SYNAPT G2-Si
  • SYNAPT G2-S
  • SYNAPT G2
  • SYNAPT HDMS

CAUSE

SYNAPT lost its tune settings.

FIX or WORKAROUND

  1. Import previous .treg file into Target Registry (if available), or retune SYNAPT in all modes and add serial number to Target Registry.
  2. Create an export file of the Target Registry settings for future use.

ADDITIONAL INFORMATION

Check Under System settings  Reflectron Grid and Pusher Offset values set at default values  is a good indication of the tune settings being lost . 
Retune these 2 settings for the best resolution  and adjust steering for best sensitivity 
Save System settings 
Save settings to PC and backup to EPC 
Default values and suggested tuning parameters are in the installation manual

id69286, MALDIG2SI, SYNAPT, SYNAPTMS, SYNG2HD, SYNG2HDMAL, SYNG2MS, SYNG2MSMAL, SYNG2SHD, SYNG2SHDML, SYNG2SIHD, SYNG2SIMS, SYNG2SMS, SYNG2SMSML

 