Can you tell what the best cone voltage is when using a cone voltage ramp? - WKB6920

Article number: 6920

ENVIRONMENT

  • ACQUITY TQD
  • MassLynx 4.1

ANSWER

No. Because the cone voltage is ramped for the duration of the scan, there is no way to pull out which cone voltage gave the best result.

ADDITIONAL INFORMATION

Setting up a series of MS methods with different cone voltages would be one way to determine the best.

Using IntelliStart sample tune and development is another option.

Manual infusion and adjustment of the cone voltage on the Tune page is a third option.

id6920, TQD, TQDIVD