Quattro Premier EPC is not booting up - WKB68770
Article number: 68770
SYMPTOMS
- EPC is not booting up
- EPC is responding to ping; response time = 1 ms
- Possible to connect to host PC (64.1.1.1) with FTP
ENVIRONMENT
- Quattro Premier
- Quattro micro API
CAUSE
The firewall is on and the FTP server is not selected in "Allowed programs and features".
FIX or WORKAROUND
Go to Control Panel > System and Security > Windows Firewall > Allowed Programs and select FTP server.
ADDITIONAL INFORMATION
Completely disabling the firewall would also solve the problem, but it may not be an acceptable solution.