No cone or capillary voltage on ZQ - WKB6788
Article number: 6788
SYMPTOMS
- No cone or capillary voltage
- Source id read-back is 10.0.
- Desolvation temperature stays at room temperature when in operate.
ENVIRONMENT
- ZQ
- Quattro Micro
CAUSE
The source enclosure O-ring is not seated correctly.
FIX or WORKAROUND
- Reseat or replace source enclosure O-ring.
- Close the source enclosure door.
- Put instrument in Operate mode.