Many low desolvation gas flow warnings in the MS Console logs - WKB65922

Article number: 65922

SYMPTOMS

  • Repeated entries in the MS console logs show "Low Desolvation Gas Flow" followed by "Instrument in Source Standby"

ENVIRONMENT

  • Xevo G2-XS
  • MassLynx
  • ACQUITY UPLC I-Class
  • PEAK nitrogen generator

CAUSE

Nitrogen generator reading low

FIX or WORKAROUND

Adjust the nitrogen pressure to read ~100 psi.

ADDITIONAL INFORMATION

 