Low sensitivity on Xevo TQ-MS with collision gas on - WKB65155

Article number: 65155

SYMPTOMS

  • Sensitivity good on MS1 with collision gas off.
  • Sensitivity low on MS2 with collision gas on.

ENVIRONMENT

  • Xevo TQ-MS
  • Xevo TQ-S

CAUSE

Incorrect ScanWave storage setting.

FIX or WORKAROUND

ScanWave storage parameters with collision gas on:

ScanWave height = 1.0 

Wave velocity = 300 

Transfer wave height = 10.0

ADDITIONAL INFORMATION

ScanWave storage parameters with collision gas off:

ScanWave height = 0 

Wave velocity = 8 

Transfer wave height = 0