Low sensitivity with Tof instrument - WKB65010
Article number: 65010
SYMPTOMS
- Poor sensitivity for sample and lockspray
- Unable to improve sensitivity with tuning
ENVIRONMENT
- Tof instrument
CAUSE
Detector voltage not optimized
FIX or WORKAROUND
Complete detector setup in IntelliStart.
ADDITIONAL INFORMATION