Low sensitivity with Tof instrument - WKB65010

Article number: 65010

SYMPTOMS

  • Poor sensitivity for sample and lockspray
  • Unable to improve sensitivity with tuning

ENVIRONMENT

  • Tof instrument

CAUSE

Detector voltage not optimized

FIX or WORKAROUND

Complete detector setup in IntelliStart.

ADDITIONAL INFORMATION