No communication with mass spectrometers in Empower - WKB6398
Article number: 6398
SYMPTOMS
- Inlet communicates fine
- Tune page does not start
- Cannot access MS console
- "Yes" status for all components
- Local Administrator works fine, general user does not
ENVIRONMENT
- Empower Network
- ZQ, SQD, TQD, SQD2, QDa
CAUSE
Insufficient priviledges for the C:\Empower\Instruments directory
FIX or WORKAROUND
- Close Empower.
- Open Windows Explorer and browse to the C:\Empower\Instruments directory.
- Right-click on this directory, select "properties", and then select the Security tab.
- Under "Group or user names" select "edit".
- Select "Add" and enter the users or group in the domain you want to access the MS. Or use "everyone" on the local PC as a user.
- Assign the rights Modify, Read & Execute, Read, Write, and List folder contents, or select "Full control".
- Reboot the computer, log in to Windows, reboot the EPC, wait five minutes, and open Empower.
ADDITIONAL INFORMATION