Low signal intensity for both sample and reference - SYNAPT - WKB63006

Article number: 63006

SYMPTOMS

Low signal intensity for both sample and reference.

ENVIRONMENT

  • SYNAPT G2
  • SYNAPT G2 MS/HDMS
  • SYNAPT G2-S MS/HDMS
  • SYNAPT G2-Si MS/HDMS
  • SYNAPT G2-S
  • SYNAPT G2-Si

CAUSE

Incorrect trap gas flow - zero flow.        

Incorrect setting on tune page or dry cylinder.    

FIX or WORKAROUND

  1. Confirm that settings are correct - manual override may be active in tune page used for acquisition. Default gas flow is 2.0.
  2. Confirm that the pressure of Argon supply is 0.5 bar.

ADDITIONAL INFORMATION

 