Low signal intensity for both sample and reference - SYNAPT - WKB63006
Article number: 63006
SYMPTOMS
Low signal intensity for both sample and reference.
ENVIRONMENT
- SYNAPT G2
- SYNAPT G2 MS/HDMS
- SYNAPT G2-S MS/HDMS
- SYNAPT G2-Si MS/HDMS
- SYNAPT G2-S
- SYNAPT G2-Si
CAUSE
Incorrect trap gas flow - zero flow.
Incorrect setting on tune page or dry cylinder.
FIX or WORKAROUND
- Confirm that settings are correct - manual override may be active in tune page used for acquisition. Default gas flow is 2.0.
- Confirm that the pressure of Argon supply is 0.5 bar.
ADDITIONAL INFORMATION
id63006, MALDIG2SI, SYNAPT, SYNAPTMS, SYNG2HD, SYNG2HDMAL, SYNG2MS, SYNG2MSMAL, SYNG2SHD, SYNG2SHDML, SYNG2SIHD, SYNG2SIMS, SYNG2SMS, SYNG2SMSML