Unable to calibrate in negative ion mode on Xevo G2-XS even when there is a good beam in the tune page or manual tune - WKB61788
Article number: 61788
SYMPTOMS
- No beam in negative ion mode when the calibration is running, but the ions are there in the tune page or manual tune
ENVIRONMENT
- MassLynx
- UNIFI
- Xevo G2-XS
- Sodium formate calibration
CAUSE
Large contaminant ion (1e7) at 126 Da, on which the system is setting the attenuation parameters
FIX or WORKAROUND
Calibrate from 150 Da and up rather than 120 Da and up.
ADDITIONAL INFORMATION