Condensation on sample cone in the Quattro Micro - WKB58817
Article number: 58817
SYMPTOMS
- Condensation on sample cone
- No droplets when flow rate is less than 0.3 mL/min
- Desolvation temperature = 450 °C
- Source temperature = 140 °C
- Desolvation gas flow rate = 250 L/h
- Starting eluent composition = 100% aqueous
ENVIRONMENT
- Quattro Micro
CAUSE
Source settings are not correct for adequate desolvation.
FIX or WORKAROUND
- Desolvation temperature recommendation: 600 to 650 °C
- Desolvation gas flow rate recommendation: 1000 L/h
- Source temperature recommendation: 150 °C