Xevo TQD shows "Source Pressure Test: Exhaust Trip" - WKB57887

Article number: 57887

SYMPTOMS

  • Source pressure test fails
  • API gas shuts off immediately after being turned on
  • No liquid in the exhaust trip

ENVIRONMENT

  • SQ Detector 2
  • Xevo TQ MS
  • Xevo TQD
  • Xevo TQ-GC
  • Xevo TQ-S
  • Xevo TQ-S micro
  • Xevo TQ-XS

CAUSE

The exhaust line coming from the top of the exhaust trap is crimped shut.

FIX or WORKAROUND

Ensure that the exhaust line is uncrimped and that it runs to a negative-draw exhaust system.

ADDITIONAL INFORMATION

 