Article number: 57887

SYMPTOMS

  • Source pressure test fails
  • API gas shuts off immediately after being turned on
  • No liquid in the exhaust trip

ENVIRONMENT

  • SQ Detector 2
  • Xevo TQ MS
  • Xevo TQD
  • Xevo TQ-GC
  • Xevo TQ-S
  • Xevo TQ-S micro
  • Xevo TQ-XS

CAUSE

The exhaust line coming from the system to the side of the exhaust trap, or the tubing from the top of the exhaust trap is crimped shut.

FIX or WORKAROUND

Ensure that the exhaust line is uncrimped and that it runs to a negative-draw exhaust system.

ADDITIONAL INFORMATION