Nitrogen gas failure - WKB55719
Article number: 55719
SYMPTOMS
- Nitrogen gas failure error message
ENVIRONMENT
- SQ Detector
- SQ Detector 2
- 3100 Mass Detector
- Xevo TQ MS
- Xevo TQD
- Xevo TQ-S
- Xevo TQ-S cronos
- Xevo TQ-S micro
- Xevo TQ-XS
CAUSE
API gas was not enabled in the tune page prior to placing mass spec into Operate.
FIX or WORKAROUND
- Turn on the API gas.
- Click on the Gas drop-down menu in the tuning page.
- Select Reset Nitrogen Gas Failure.