ACQUITY QDa Detector gives spikey chromatagraphic peaks that drop to baseline - WKB55479
Article number: 55479
SYMPTOMS
- Spikey MS chromatographic peaks:
- Whereas the following is expected/normal:
ENVIRONMENT
- ACQUITY QDa Detector
- MassLynx
- Autopurification
CAUSE
Damaged probe
FIX or WORKAROUND
Replace the ESI PROBE ASSEMBLY 250 mm (700009641).
ADDITIONAL INFORMATION
id55479, Electrospray, QDA