Cannot start vacuum system on the SQ Detector 2 - WKB55409
Article number: 55409
SYMPTOMS
- SQ Detector 2 (SQD2) vacuum system found off
- Cannot restart vacuum system
- Rebooting system does not resolve
- Found: MS status "Instrument Present"
- Found: Operate LED on SQD2 is RED
ENVIRONMENT
- SQ Detector 2
- Xevo TQD
CAUSE
Partial communication issue between MassLynx PC and embedded PC (EPC)
FIX or WORKAROUND
- Close MassLynx.
- Reboot the MassLynx PC. Log in to Windows but do not start MassLynx.
- Power-off the SQD2. Wait one minute.
- Power-on the SQD2. Wait for the Operate LED to go from RED to OFF.
- When the Operate LED turns off, launch MassLynx.
- Verify that the MS status is "Not Scanning".
- Open the Tune page. From the menus, go to Vacuum > Pump.
ADDITIONAL INFORMATION
id55409, A-SQ, comms, MLYNX, MLYNXV41, SQD2, SUPMM