MS peak area increasing injection to injection over run - WKB52784
Article number: 52784
SYMPTOMS
- MS peak area increasing injection to injection over run.
- Method dependent.
- All ACQUITY diagnostics pass.
- Refreshing sample, mobile phase, and column did not resolve issue.
ENVIRONMENT
- XEVO TQ MS
- MassLynx
- Xevo TQD
- Xevo TQ-S
CAUSE
Charging effect caused by a dirty ion block and guide.
FIX or WORKAROUND
- Clean ion block and guide following the appropriate Overview and Maintenance Guide.
ADDITIONAL INFORMATION
A dirty source enclosure can also cause this charging effect, so if cleaning the ion block, clean the source enclosure as well.
id52784, eluent, MLYNX, MLYNXV41, SUPMM, UPBINARY, XEVOTQ, XEVOTQA, XEVOTQD, XEVOTQDIVD, XEVOTQIVD, XEVOTQS, XEVOTQSIVD, XEVTQSMIVD