How do I drain the oil from the SOGEVAC SV65 BI rotary pump? - WKB52355

Article number: 52355

OBJECTIVE or GOAL

Drain the oil from the SOGEVAC SV65 BI rotary pump.    

ENVIRONMENT

  • SOGEVAC SV65 BI rotary pump
  • Xevo TQ-S

PROCEDURE

  1. Vent the mass spectrometer.
  2. Remove the fill port cap from the pump.
  3. Remove the drain port plug from the pump.
  4. Install the drain port tube (has a brass fitting that matches that of the drain port plug).