Low sensitivty on a Xevo TQ MS - WKB51476

Article number: 51476

SYMPTOMS

  • Low sensitivity when running analytical methods
  • Low sensitivity with setup solution

ENVIRONMENT

  • Xevo TQ MS

CAUSE

Dirty source and hexapole

FIX or WORKAROUND

  1. Clean the source components.
  2. Clean the hexapole.

ADDITIONAL INFORMATION

Find instructions for cleaning these components in the Xevo TQ MS Operator’s Overview and Maintenance Guide.