Skip to main content
Waters

Sudden loss of resolution on SYNAPT G2-Si - WKB49063

Article number: 49063

SYMPTOMS

  • Sudden loss of resolution found in runs; earlier runs that day are fine
  • Resolution in positive resolution mode found to be 3k; in high resolution mode the reading was 6k

ENVIRONMENT

  • MassLynx
  • SYNAPT G2-Si

CAUSE

Unknown

FIX or WORKAROUND

  1. Open the tune page and go to the Setup menu.
  2. Select "System View".
  3. Enter the password:  access
  4. Next, go to the File menu.
  5. Choose "Load System Settings (from PC)".
  6. Determine if resolution returns by infusing a sample.
  7. When satisfied, go back to the Setup menu and select for System View to exit out of this view. (Note: You will be prompted to do a System Save. Make an appropriate choice.)

ADDITIONAL INFORMATION

Not able to find a solution? Click here to request help.