Xevo TQXXX gives "Desolvation temperature failed to settle" error - WKB47385
Article number: 47385
SYMPTOMS
- "Desolvation temperature failed to settle" error
- Rebooting the system does not remove the error
ENVIRONMENT
- Xevo TQD
- SQD2
- Xevo TQ-S micro
- Xevo G2-XS
- Synapt G2Si
CAUSE
Inconsistent nitrogen flow from the dewar or nitrogen generator
FIX or WORKAROUND
Confirm a steady nitrogen supply from the dewar or generator:
- Verify that the supply lines are fitted correctly.
- Verify that the dewar is not empty.
- Verify that the nitrogen generator is working properly.
ADDITIONAL INFORMATION