Xevo TQD gives "Desolvation temperature failed to settle" error - WKB47385

Article number: 47385

SYMPTOMS

  • "Desolvation temperature failed to settle" error
  • Rebooting the system does not remove the error

ENVIRONMENT

  • Xevo TQD
  • SQD2
  • Xevo TQ-S micro
  • Xevo G2-XS
  • Synapt G2Si

CAUSE

Inconsistent nitrogen flow from the dewar or nitrogen generator

FIX or WORKAROUND

Confirm a steady nitrogen supply from the dewar or generator:

  • Verify that the supply lines are fitted correctly.
  • Verify that the dewar is not empty.
  • Verify that the nitrogen generator is working properly.

ADDITIONAL INFORMATION

 