Skip to main content
Waters

Xevo TQXXX gives "Desolvation temperature failed to settle" error - WKB47385

Article number: 47385

SYMPTOMS

  • "Desolvation temperature failed to settle" error
  • Rebooting the system does not remove the error

ENVIRONMENT

  • Xevo TQD
  • SQD2
  • Xevo TQ-S micro
  • Xevo G2-XS
  • Synapt G2Si

CAUSE

Inconsistent nitrogen flow from the dewar or nitrogen generator

FIX or WORKAROUND

Confirm a steady nitrogen supply from the dewar or generator:

  • Verify that the supply lines are fitted correctly.
  • Verify that the dewar is not empty.
  • Verify that the nitrogen generator is working properly.

ADDITIONAL INFORMATION

 

id47385, A-SQ, MALDIG2SI, SQD2, SYNG2SIHD, XEVOG2XSTF, XEVOTQD, XEVOTQDIVD, XEVOTQSMIC

Not able to find a solution? Click here to request help.