Cannot put mass spectrometer into Standby or turn off the API gas from the tune page or the ACQUITY MS console - WKB46204
Article number: 46204
SYMPTOMS
- Cannot put mass spectrometer into Standby; button is unresponsive
- Cannot turn off API gas; button is unresponsive
- Found: MS status at bottom of sample list page is "Instrument Present"
ENVIRONMENT
- All Ethernet-based mass spectrometers controlled by MassLynx
CAUSE
No communication between MassLynx PC and embedded PC (EPC)
FIX or WORKAROUND
- Close MassLynx
- Reboot the MassLynx PC.
- Log in to Windows but do not start MassLynx.
- Push the reset button on the mass spectrometer to reboot the EPC.
- Wait until EPC reboots. Can take up to 3 minutes.
- Open MassLynx.