No MS signal on data acquired in MassLynx - WKB44688

Article number: 44688

SYMPTOMS

  • Data acquired has no peaks for the MS traces, yet there is UV signal
  • Earlier runs show signal for both MS and UV

ENVIRONMENT

  • Xevo G2-S

CAUSE

Leak from onboard fluidics valve

FIX or WORKAROUND

Tighten fittings on the fluidics valves.

ADDITIONAL INFORMATION

 