Tips for how to opimize IonKey signal in negative mode - WKB44582
Article number: 44582
OBJECTIVE or GOAL
To enhance negative mode signal strength with IonKey.
ENVIRONMENT
- IonKey negative mode
- Mass Spectrometry
PROCEDURE
- Generally, capillary voltage optimizes lower in negative mode than it does in positive mode. For the best results, tune low to high.
- Squirt the emitter tip with methanol to clean off foreign debris. Foreign debris in negative mode is much more detrimental to sensitivity than in positive mode.
- Be cautious if the instrument is also used for positive mode acquisitions. Any lingering acid from a positive mode acquisition can also be a significant ion suppressor.
- Keep in mind that negative mode acquisitions have less signal strength than positive mode acquisitions, but there is also less noise.
ADDITIONAL INFORMATION
id44582, IONKEY