Error: Sample tune check failed in the ACQUITY TQD - WKB4457
Article number: 4457
SYMPTOMS
- Console displays "Sample tune check failed" error
ENVIRONMENT
- ACQUITY TQD
- Empower
CAUSE
The sample tune has expired; when it runs, it fails because the check system is not set up.
FIX or WORKAROUND
Clear the "sample tune expired" selection in the IntelliStart configuration:
- Open the console.
- Select TQD in the left pane.
- Select Configure > IntelliStart configuration.
- Clear Positive and negative sample tune expired.
- Click OK.