After PM Mass spec begins acquiring data as soon as sample list is started - WKB4153
Article number: 4153
SYMPTOMS
- PM recently performed
- Mass spec begins acquiring data as soon as sample list is started.
- Mass spec does not wait for LC to inject
ENVIRONMENT
- All MassLynx controlled Mass Spectrometers
CAUSE
Contact closure (event 1) is deconfigured in inlet method
FIX or WORKAROUND
- In Inlet Method, select "Event in 1".
ADDITIONAL INFORMATION
Engineers often clear "Event in 1" for instrument testing after a PM.
id4153, MLYNX, MLYNXV41, SUPMM