After PM Mass spec begins acquiring data as soon as sample list is started - WKB4153
Article number: 4153
SYMPTOMS
- PM recently performed
- Mass spec begins acquiring data as soon as sample list is started.
- Mass spec does not wait for LC to inject
ENVIRONMENT
- All MassLynx controlled Mass Spectrometers
CAUSE
Contact closure (event 1) is deconfigured in inlet method
FIX or WORKAROUND
- In Inlet Method, select "Event in 1".
ADDITIONAL INFORMATION
Engineers often clear "Event in 1" for instrument testing after a PM.
id4153, A-SQ, A-SQ, CYCIMS, MALDIG2SI, MLYNX, MLYNXV41, MRT, Q3100MSD, QDA, RADIAN, RDA, SQD, SQD2, SUPMM, SYNAPT, SYNAPTMS, SYNAPTXS, SYNG2HD, SYNG2HDMAL, SYNG2MS, SYNG2MSMAL, SYNG2SHD, SYNG2SHDML, SYNG2SIHD, SYNG2SIMS, SYNG2SMS, SYNG2SMSML, TQD, TQDIVD, XEVOG2QTOF, XEVOG2SQTF, XEVOG2STOF, XEVOG2TOF, XEVOG2XSQT, XEVOG2XSTF, XEVOG3QTOF, XEVOQTOF, XEVOTQ, XEVOTQA, XEVOTQA, XEVOTQD, XEVOTQDIVD, XEVOTQGC, XEVOTQIVD, XEVOTQS, XEVOTQSCRO, XEVOTQSIVD, XEVOTQSMIC, XEVOTQXS, XEVTQSMIVD, XEVTQXSIVD