Skip to main content
Waters

After PM Mass spec begins acquiring data as soon as sample list is started - WKB4153

Article number: 4153

SYMPTOMS

  • PM recently performed
  • Mass spec begins acquiring data as soon as sample list is started. 
  • Mass spec does not wait for LC to inject

ENVIRONMENT

  • All MassLynx controlled Mass Spectrometers

CAUSE

Contact closure (event 1) is deconfigured in inlet method

FIX or WORKAROUND

  1. In Inlet Method, select "Event in 1".

ADDITIONAL INFORMATION

Engineers often clear "Event in 1" for instrument testing after a PM.

Not able to find a solution? Click here to request help.