ACQUITY Standard QDa/RADIAN ASAP turbo timeout - WKB3794

Article number: 3794

SYMPTOMS

  • ACQUITY QDa/RADIAN ASAP crash vented during acquisition.
  • ACQUITY QDa/RADIAN ASAP failing to pump back down, only reaches 40% turbo speed.
  • Turbo pump power very high (>70W)
  • Turbo speed does not reach 100%
  • Cannot pump down

ENVIRONMENT

  • ACQUITY QDa Mass Detector (Standard)
  • RADIAN ASAP Direct Mass Detector (Standard)

CAUSE

Diaphragm pump not performing at optimum efficiency.

FIX or WORKAROUND

  1. Rebuild the diaphragm pump using the service kit #700009680.
  2. Contact ESTC refurb program for replacement.
  3. If new installation consider replacing the diaphragm pump with new #700009640.

Pump down instrument.

Verify that turbo pump power is now within normal range (around 45W)

ADDITIONAL INFORMATION