How to vent and shut down a quadrupole MS system for a planned power outage - WKB32587

Article number: 32587

OBJECTIVE or GOAL

Safely shut down the LCMS system in preparation for a planned power outage.

ENVIRONMENT

  • MassLynx
  • TQD
  • SQD
  • Xevo TQD
  • SQD2
  • Xevo TQ
  • Xevo TQ-S

PROCEDURE

  1. Turn off LC flow into the source.
  2. Put the MS into Standby mode.
  3. Turn off all gas flows.
  4. Vent the instrument (from the Tune Page / Vacuum pull-down menu, choose 'Vent'). In Diagnostics tab monitor the Turbos speed and await until those drops to about 2%.
  5. Once the system has vented, close MassLynx.
  6. Power-off the MS.
  7. Unplug the main power cable to the MS.
  8. Power-off the LC modules using power switch on the front of each module.
  9. Shut down the PC.