No LockSpray reference signal - WKB32556
Article number: 32556
SYMPTOMS
- No signal from LockSpray reference probe
- Liquid flow is present on fused silica line to probe
- Sample spray visible
- Detector noise visible
ENVIRONMENT
- SYNAPT G2-Si
- SYNAPT G2-S
- SYNAPT G2
- Xevo G2-XS
- Xevo G2-S
- Xevo G2
- nanoLockSpray
CAUSE
Incorrectly built reference sprayer
FIX or WORKAROUND
Rebuild the reference sprayer.
ADDITIONAL INFORMATION
Refer to the Maintenance Procedures section of the appropriate instrument Overview and Maintenance Guide for guidance on rebuilding the reference sprayer.
