How to replace the fluidics tubing that goes to the A, B, or C bottles on a SYNAPT G2 - WKB29496
Article number: 29496
OBJECTIVE or GOAL
Replace the orange tubing that goes from the sample valve (middle valve) to the A, B, or C bottles on the fluidics.
ENVIRONMENT
- SYNAPT G2
- SYNAPT G2 MS/HDMS
- SYNAPT G2-S MS/HDMS
- SYNAPT G2-Si MS/HDMS
- SYNAPT G2-S
- SYNAPT G2-Si
- Vion IMS QTof
- Xevo QTof
- Xevo Tof
- Xevo G2 QTof
- Xevo G2 Tof
- Xevo G2-S
- Xevo G2-S QTof
- Xevo G2-S Tof
- Xevo G2-XS
- Xevo G2-XS QTof
- Xevo G2-XS Tof
- SQ Detector 2
- Xevo TQ MS
- Xevo TQD
- Xevo TQ-S
- Xevo TQ-S micro
- Xevo TQ-XS
PROCEDURE
- Open the door or doors by the fluidics (top and bottom).
- Loosen and remove the fittings above the bottles.
- On the sample fluidics valve (middle), loosen the fitting and remove the tubing that corresponds to the bottle the line to be replaced goes to (A = port 4, B = port 5, C = port 6).
- Using tape, connect the new tube (from kit 700008827) to the old tube.
- Pull the old tubing or push the new tubing through the panel to feed the new tubing into place.
- Disconnect the old tubing and trim off the end with tape on it (using a proper tubing cutter).
- Tighten the remaining fittings and close the door.
- Put some IPA into a bottle in the position where the tubing was replaced.
- Purge 4 or 5 times.
ADDITIONAL INFORMATION
id29496, A-SQ, isopropanol, MALDIG2SI, SQD2, SYNG2HD, SYNG2HDMAL, SYNG2MS, SYNG2MSMAL, SYNG2SHD, SYNG2SHDML, SYNG2SIHD, SYNG2SIMS, SYNG2SMS, SYNG2SMSML, XEVOG2QTOF, XEVOG2SQTF, XEVOG2STOF, XEVOG2TOF, XEVOG2XSQT, XEVOG2XSTF, XEVOQTOF, XEVOTQ, XEVOTQA, XEVOTQD, XEVOTQDIVD, XEVOTQIVD, XEVOTQS, XEVOTQSIVD, XEVOTQSMIC, XEVOTQXS, XEVTQSMIVD, XEVTQXSIVD