Skip to main content
Waters

QDa fails source pressure test - WKB287546

Article number: 287546

SYMPTOMS

  • QDa fails source pressure test

ENVIRONMENT

  • QDa

CAUSE

Poor seal between source enclosure and seal/exhaust tubing not seated correctly

FIX or WORKAROUND

  1. Reseat the source enclosure.
  2. Reseat the exhaust tubing.
  3. Rerun the source pressure test.

ADDITIONAL INFORMATION

 

id287546, QDA

Not able to find a solution? Click here to request help.