QDa fails source pressure test - WKB287546

Article number: 287546

SYMPTOMS

  • QDa fails source pressure test

ENVIRONMENT

  • QDa

CAUSE

Poor seal between source enclosure and seal/exhaust tubing not seated correctly

FIX or WORKAROUND

  1. Reseat the source enclosure.
  2. Reseat the exhaust tubing.
  3. Rerun the source pressure test.