Skip to main content
Waters

No communications with new EPC; no instrument type or serial number - WKB285365

Article number: 285365

SYMPTOMS

  • EPC replaced for communications issue, but no communications on reboot
  • Hyperterminal/puTTy log is normal and successful
  • No instrument type or serial number shows in Waters DHCP Server Wizard
  • Checked with laptop and no communications 

ENVIRONMENT

  • MassLynx
  • Xevo TQ-S micro

CAUSE

Setting on PC is restricting communications, and PC (laptop) is unsupported.

FIX or WORKAROUND

Reimage the PC or try a new PC.

ADDITIONAL INFORMATION

 

Not able to find a solution? Click here to request help.