Source pressure test failure - WKB279255

Article number: 279255

SYMPTOMS

  • Source pressure test failure
  • Plugging the John Guest valve allows system to pass source pressure test

ENVIRONMENT

  • Xevo G2-S Tof

CAUSE

Faulty trap bottle

FIX or WORKAROUND

Replace the John Guest valve (6063338) and/or the trap bottle (700004606).

ADDITIONAL INFORMATION

 