How to reboot the EPC on Xevo instruments using Telnet - WKB274059
Article number: 274059
OBJECTIVE or GOAL
Reboot the EPC (embedded PC) on a Xevo system.
(This is required after restarting the acquisition PC and/or MassLynx.)
ENVIRONMENT
- Xevo TQ Absolute
- Xevo TQ MS
- Xevo TQD
- Xevo TQ-S
- Xevo TQ-S cronos
- Xevo TQ-S micro
- Xevo TQ-XS
- Xevo QTof
- Xevo G2 QTof
- Xevo G2 Tof
- Xevo G2-S
- Xevo G2-S QTof
- Xevo G2-S Tof
- Xevo G2-XS
- Xevo G2-XS QTof
- Xevo G2-XS Tof
- Xevo G3 QTof
PROCEDURE
- Using the start menu, type "cmd".
- Open the "Command Prompt" app.
- In the command prompt app, type "telnet epc". This gives direct command control of the embedded PC in the MS via telnet.
- Type "reboot" in the window next to the "->". This restarts the embedded PC on the MS.
- After five minutes, you can log back in to MassLynx.
ADDITIONAL INFORMATION
This procedure should work on most Waters mass spectrometers.