SYMPTOMS

  • Reference fluidics system error in console log
  • Divert valve flipped to waste

ENVIRONMENT

  • Xevo G2-XS
  • MassLynx
  • SYNAPT G2
  • SYNAPT G2 MS/HDMS
  • SYNAPT G2-S MS/HDMS
  • SYNAPT G2-Si MS/HDMS
  • SYNAPT G2-S
  • SYNAPT G2-Si
  • Xevo QTof
  • Xevo Tof
  • Xevo G2 QTof
  • Xevo G2 Tof
  • Xevo G2-S
  • Xevo G2-S QTof
  • Xevo G2-S Tof
  • Xevo G2-XS
  • Xevo G2-XS QTof
  • Xevo G2-XS Tof

CAUSE

Unknown

FIX or WORKAROUND

  1. On the Tune page, go to the Setup drop-down menu and select System View.
  2. Enter the password: access
  3. Go to the Fluidics tab.
  4. Click the reset button (not the purge button).

fluidics reset.PNG