Frequent Lockmass check errors when using dynamic Lockmass on a BioAccord or ACQUITY RDA - WKB268624

Article number: 268624

SYMPTOMS

  • Frequent Lockmass check errors when using dynamic Lockmass on a BioAccord or ACQUITY RDA

ENVIRONMENT

  • BioAccord
  • ACQUITY RDA
  • waters_connect 3.1.1 or preceding 

CAUSE

Over-attenuation of leucine enkephalin due to the attenuation of caffeine or pentane sulphonic acid

FIX or WORKAROUND

Do not use the dynamic lockmass option in the analysis method.

ADDITIONAL INFORMATION

A fix for this issue in the RDA driver will be released with waters_connect 3.2.0.

id268624, RDA