Frequent Lockmass check errors when using dynamic Lockmass on a BioAccord or ACQUITY RDA - WKB268624
Article number: 268624
SYMPTOMS
- Frequent Lockmass check errors when using dynamic Lockmass on a BioAccord or ACQUITY RDA
ENVIRONMENT
- BioAccord
- ACQUITY RDA
- waters_connect 3.1.1 or preceding
CAUSE
Over-attenuation of leucine enkephalin due to the attenuation of caffeine or pentane sulphonic acid
FIX or WORKAROUND
Do not use the dynamic lockmass option in the analysis method.
ADDITIONAL INFORMATION
A fix for this issue in the RDA driver will be released with waters_connect 3.2.0.
id268624, RDA