High response for C5DC in neonatal screening - WKB268555
Article number: 268555
SYMPTOMS
- High response for C5DC in neonatal screening patient samples (MRM transition: 276.2>85.0)
- Other analytes in neonatal screening in range
- QC samples from PerkinElmer NeoBase kit look good
- Problem appears after routine service on nitrogen generator
ENVIRONMENT
- Xevo TQ-S micro
CAUSE
Contamination from the nitrogen generator
FIX or WORKAROUND
Ask PEAK Scientific to service the nitrogen generator.
ADDITIONAL INFORMATION