RDa nitrogen gas below limit, faulty PEAK N2 GENIUS SQ24 generator - WKB266066
Article number: 266066
SYMPTOMS
- Nitrogen gas below limit
ENVIRONMENT
- RDa
- BioAccord
- PEAK N2 GENIUS SQ24
CAUSE
Faulty N2 generator. Nitrogen gas below limit. Nitrogen must be between 6.5 and 7.5 bar.
FIX or WORKAROUND
Peak identified a solenoid valve that was stuck open with a small piece of another valve and replaced both parts.
ADDITIONAL INFORMATION
id266066, RDA