QDa Voltage Readback failure - WKB261154
Article number: 261154
SYMPTOMS
- Voltage Readback error seen in Console
- Power cycle of instrument alone does not clear the error
ENVIRONMENT
- ACQUITY QDa Mass Detector
CAUSE
Critical readback check report shows one or more of the following checks fail:
Ion Guide 1 Offset (Lens 4) Readback
StepWave Head Gradient (Lens 5) Readback
Step Wave Offset (Lens 6) Readback
StepWave Body Gradient (Lens 7) Readback
Source Offset (Lens 8) Readback
Sample Cone Voltage (Lens 3) Readback
Impactor Readback
Dynode Voltage Readback
Phosphor Voltage Readback
SnowBoy RF Amplitude Readback
FIX or WORKAROUND
- Power-off the QDa.
- Close the software.
- Reboot the PC.
- Log in to Windows.
- Power-on the QDa.
- Wait 2-3 minutes.
- Open the software and confirm communication connection.
ADDITIONAL INFORMATION
The Critical Readback report can be found in the console when using MassLynx or in the following directory when using Empower: :\Empower\Instruments\Log\MS Reports.
id261154, comms, QDA