No communications with the SQ Detector 2 after MassLynx workstation is put into sleep mode - WKB259039
Article number: 259039
SYMPTOMS
- MassLynx workstation put into sleep mode
- In MassLynx, no communication with the SQD2; ms status = no instrument
- SQD2 does not show in the MS Console
- Communications with the ACQUITY UPLC components are OK
- Rebooting the EPC via the reset button does not resolve
ENVIRONMENT
- SQ Detector 2
- MassLynx
CAUSE
Corrupted IP address
FIX or WORKAROUND
- Close MassLynx and the ACQUITY MS Console.
- Open the Waters DHCP Server Configuration.
- Remove the IP address for the SQD2.
- Restart the MassLynx workstation.
- Open the Waters DHCP Server Configuration.
- Push the reset button on the SQD2. Wait for the chime.
- Open MassLynx and confirm that Not Scanning status can be seen in the MS Console.