Oil leakage on Edwards nRV14i rotary pump - RDa - WKB253774

Article number: 253774

SYMPTOMS

  • Slight oil leakage on top of Edwards pump

ENVIRONMENT

  • Edwards nRV14i rotary pump

CAUSE

The O-ring between the oil demister and the pump outlet is twisted.

FIX or WORKAROUND

  1. Remove the O-ring.
  2. Untwist the ring and put it back properly.
  3. If the ring is damaged, replace it (part 6071201 (aluminium) or part 6071221 (polymer).