What is the optimal temperature range for a Xevo TQ-S? - WKB250675
Article number: 250675
ENVIRONMENT
- Xevo TQ-S
ANSWER
The ambient temperature range required for normal operation is 15 to 28 °C (59 to 82 °F).
ADDITIONAL INFORMATION
Temperature stability must be better than 2 °C or 3.5 °F peak to peak in 1.5 hours.
This and other information can be found in the Waters Xevo TQ-S Site Preparation Guide (document 715002440).