ACQUITY QDa Mass Detector source pressure test failure - WKB250587

Article number: 250587

SYMPTOMS

  • QDa source pressure test failure 
  • If overriding the SPT, the QDa runs normally
  • QDa "source exhaust pressure leak detected" error in console
  • QDa source pressure test fails, giving a source pressure monitoring leak warning

ENVIRONMENT

  • ACQUITY QDa Mass Detector

CAUSE

Failed bottle trap assembly

FIX or WORKAROUND

Replace the bottle trap assembly (part 700009635).

ADDITIONAL INFORMATION

 

id250587, QDA