The TUV is missing from the MS console - WKB24985
Article number: 24985
SYMPTOMS
- The TUV is not present in the MS console or the ACQUITY console
ENVIRONMENT
- MassLynx
- ACQUITY UPLC TUV
- Empower
- ACQUITY UPLC PDA
- ACQUITY UPLC FLR
CAUSE
Unknown
FIX or WORKAROUND
- Close MassLynx or Empower and MS or ACQUITY console, whichever apply.
- Power-off the TUV or the detector having issues.
- Open the Waters DHCP Server configuration wizard.exe in C:\Program files x86\Waters Instruments\Waters DHCP Server Configuration.exe (if using Empower, then look in the C:\Empower\Instruments folder).
- Remove the detector from the list of components found in the DHCP Server Configuration window.
- Power-on the detector and monitor the DHCP server window until the detector repopulates in all four columns.
- Launch the software and the console and verify that the detector is now present.
ADDITIONAL INFORMATION
id24985, A-30FL, A-30PDA, A-30UV, EMP2LIC, EMP2OPT, EMP2SW, EMP3GC, EMP3LIC, EMP3OPT, EMP3SW, EMPGC, EMPGPC, EMPLIC, EMPOWER2, EMPOWER3, EMPSW, MLYNX, MLYNXV41, SUP, SUPMM, UPAPDA, UPARPPDA, UPARPTUV, UPC2PDA, UPFLRARC, UPFLRDET, UPPDA, UPPDAARC, UPPDAARCB, UPPDA-E, UPPDA-L, UPPDALTC, UPPDATC, UPPFLR, UPPPDA, UPPTUV, UPTUV, UPTUVARC, UPTUVARCB, UPTUV-E, UPTUVTC