How to run detector setup on a Xevo G2-XS QTof - WKB24981
Article number: 24981
OBJECTIVE or GOAL
Run detector setup.
ENVIRONMENT
- SYNAPT G2
- SYNAPT G2 MS/HDMS
- SYNAPT G2-S MS/HDMS
- SYNAPT G2-Si MS/HDMS
- SYNAPT G2-S
- SYNAPT G2-Si
- SYNAPT XS
- Xevo G2 QTof
- Xevo G2 Tof
- Xevo G2-S
- Xevo G2-S QTof
- Xevo G2-S Tof
- Xevo G2-XS
- Xevo G2-XS QTof
- Xevo G2-XS Tof
PROCEDURE
- Select the mode (Res/Sens) you are mostly using in your analysis.
- Open the console and go to the IntelliStart section under the MS.
- Select Detector Setup and click Start (on the right).
- On the tune page, infuse leucine enkephalin at a flow rate and capillary voltage to give a stable beam (5e5–1e6 cps in Positive Ion mode). Adjust the counts by controlling the capillary voltage, cone voltage, concentration, or flow rate.
- Work through the detector setup wizard, selecting the tune page for fluidics settings and flow rate.