Skip to main content
Waters

How to run detector setup on a Xevo G2-XS QTof - WKB24981

Article number: 24981

 

OBJECTIVE or GOAL

Run detector setup.

ENVIRONMENT

  • SYNAPT G2
  • SYNAPT G2 MS/HDMS
  • SYNAPT G2-S MS/HDMS
  • SYNAPT G2-Si MS/HDMS
  • SYNAPT G2-S
  • SYNAPT G2-Si
  • SYNAPT XS
  • Xevo G2 QTof
  • Xevo G2 Tof
  • Xevo G2-S
  • Xevo G2-S QTof
  • Xevo G2-S Tof
  • Xevo G2-XS
  • Xevo G2-XS QTof
  • Xevo G2-XS Tof

PROCEDURE

  1. Open the console and go to the IntelliStart section under the MS.

Intellistart_configure_mode.PNG

  1. Select Detector Setup and click Start (on the right).
  2. On the tune page, infuse leucine enkephalin at a flow rate and capillary voltage to give a stable beam (5e5–1e6 cps in Positive Ion mode). Adjust the counts by controlling the capillary voltage, cone voltage, concentration, or flow rate.
  3. Work through the detector setup wizard, selecting the tune page for fluidics settings and flow rate.

ADDITIONAL INFORMATION

 

id24981, MALDIG2SI, SYNAPTXS, SYNG2HD, SYNG2HDMAL, SYNG2MS, SYNG2MSMAL, SYNG2SHD, SYNG2SHDML, SYNG2SIHD, SYNG2SIMS, SYNG2SMS, SYNG2SMSML, XEVOG2QTOF, XEVOG2SQTF, XEVOG2STOF, XEVOG2TOF, XEVOG2XSTF

 

Not able to find a solution? Click here to request help.